Langer E1 set | PCB Immunity Development System

Approx. 0-1500 V pulses | 2 ns rise / 10 ns tail | 4 H-field + 5 E-field sources | 10 Mbps optical sensor

E1 combines the SGZ 21 burst generator with magnetic and electric field sources for finding PCB weaknesses associated with EFT/burst and ESD. Apply a localized disturbance, observe the failure, then compare the result after a grounding, shielding or circuit change. The compact bench system includes an S21 optical sensor, an MS 02 magnetic probe, accessories, power supply, case and manual.

E1 set - Datasheet

SPECIFICATIONS

E1 set - PCB Immunity Development System

E1 combines the SGZ 21 burst generator with magnetic and electric field sources for finding PCB weaknesses associated with EFT/burst and ESD. Apply a localized disturbance, observe the failure, then compare the result after a grounding, shielding or circuit change. The compact bench system includes an S21 optical sensor, an MS 02 magnetic probe, accessories, power supply, case and manual.

E1 set, Immunity Development System

Features and application benefits

FeatureEngineering benefit / use
Included coupling tools Four magnetic field sources and five electric field sources; MS 02 is a sensing probe, not another field generator.
Pulse values Approximate generator values are 0-1500 V peak, 2 ns rise time and 10 ns tail time. They are not DUT pin-voltage guarantees.
Optical monitoring S21 supports DC to 10 Mbps; the SGZ 21 optical input is separately rated at 5 MHz with a 100 ns minimum pulse width.

Technical specifications

ParameterSpecification
S21 optical sensor / Transmission range DC...10 Mbps
S21 optical sensor / Optical fibre connector 2.2 mm diameter
S21 optical sensor / Supply voltage (3-5) V
S21 optical sensor / Current input 10 mA
SGZ 21 burst Generator / Pulse parameter / Rise time ca. 2 ns
SGZ 21 burst Generator / Pulse parameter / Tail time ca. 10 ns
SGZ 21 burst Generator / Pulse parameter / Peak values ca. 0...1500 V
SGZ 21 burst Generator / Optical input / Optical fiber 2.2 mm
SGZ 21 burst Generator / Optical input / Max. frequency 5 MHz
SGZ 21 burst Generator / Optical input / Min. pulse width 100 ns
SGZ 21 burst Generator / Optical input / Supply voltage 12 V / 200 mA
SGZ 21 burst Generator / Optical input / Sizes (L x W x H) (154 x 100 x 62) mm

Included components

PhotoQty.ItemFunction / key specifications
SGZ 21, Burst Generator 1x SGZ 21 Burst Generator
Pulse parameter / Rise time: ca. 2 ns
Pulse parameter / Tail time: ca. 10 ns
Pulse parameter / Peak values: ca. 0... 1500 V
Optical input / Optical fiber: 2.2 mm
S21, Optical Sensor (10 Mbps) 1x S21 Optical Sensor (10 Mbps)
Transmission range: DC ... 10 Mbps
Optical fibre connector: 2.2 mm diameter
Supply voltage: (3 -5) V
Current input: 10 mA
BS 02, Magnetic Field Source 1x BS 02 Magnetic Field Source
BS 04DB, Magnetic Field Source 1x BS 04DB Magnetic Field Source
BS 05D, Magnetic Field Source 1x BS 05D Magnetic Field Source
BS 05DU, Magnetic Field Source 1x BS 05DU Magnetic Field Source
ES 00, E-Field Source 1x ES 00 E-Field Source
ES 01, E-Field Source 1x ES 01 E-Field Source
ES 02, E-Field Source 1x ES 02 E-Field Source
ES 05D, E-Field Source 1x ES 05D E-Field Source
ES 08D, E-Field Source 1x ES 08D E-Field Source
MS 02, Magnetic Field Probe 1x MS 02 Magnetic Field Probe
  1x E1 acc Accessories
  1x NT FRI EU Power Supply Unit
  1x E1 case System Case
  1x E1 m E1 Set User Manual

Included component specifications

ComponentParameterSpecification
SGZ 21 Pulse parameter / Rise time ca. 2 ns
SGZ 21 Pulse parameter / Tail time ca. 10 ns
SGZ 21 Pulse parameter / Peak values ca. 0... 1500 V
SGZ 21 Optical input / Optical fiber 2.2 mm
SGZ 21 Optical input / Max. frequency 5 MHz
SGZ 21 Optical input / Min. pulse width 100 ns
SGZ 21 Counter / Display 6 digit
SGZ 21 Counter / Peak time 1 s
SGZ 21 Counter / Supply voltage 12 V / 200 mA
SGZ 21 Counter / Sizes (L x W x H) (154 x 100 x 62) mm
S21 Transmission range DC ... 10 Mbps
S21 Optical fibre connector 2.2 mm diameter
S21 Supply voltage (3 -5) V
S21 Current input 10 mA
S21 Sizes (L x W x H) (34 x 10 x 6) mm

Application and kit views

ImageWhat it shows
Application with BS 04 DB Application with BS 04 DB
Scope of delivery E1 Scope of delivery E1
Exploded view drawing E1 Exploded view drawing E1
Application with MSA 02 magnetic field probe from the S2 set Application with MSA 02 magnetic field probe from the S2 set
Measuring principles Measuring principles

System requirements and ordering notes

SGZ 21 included; 12 V / 200 mA. A powered DUT and a way to observe its functional response are required. The tools support engineering diagnosis; they do not replace the complete standardized compliance test setup.

Recommended products and accessories

PhotoProductSpecifications / purposeOrdering
S2 set, Magnetic Field Probes for E1 S2 set Active and passive H-field sensing; MSA 02 with 3 heads, MS 101 and MS 102U. One AAA battery is required for MSA 02. Optional set; not included

Datasheets

DocumentPDF
E1 set - Datasheet View PDF

Practical questions

QuestionAnswer
What is this product best suited to? Start with a complete bench toolkit for isolating PCB susceptibility.
What supporting equipment is needed? SGZ 21 included; 12 V / 200 mA. A powered DUT and a way to observe its functional response are required. The tools support engineering diagnosis; they do not replace the complete standardized compliance test setup.
Does this establish compliance with IEC 61000-4-4 or IEC 61000-4-2? These are development and diagnostic tools. Relating a local disturbance to an EFT/burst or ESD failure helps troubleshooting; formal compliance requires the applicable standardized generator, coupling method, test setup and procedure.

Product videos

PCB immunity selection guide

Compare generation, coupling and detection tools. The kit pictures identify the complete sets. Product links below use the planned Absolute EMC product pages.

KitProduct / functionKey distinctionRequired pulse source / power
E1 set, Immunity Development System E1 set
Integrated generation and local E/H-field troubleshooting
Approx. 0-1500 V pulses | 2 ns rise / 10 ns tail | 4 H-field + 5 E-field sources | 10 Mbps optical sensor SGZ 21 included; 12 V / 200 mA
E2 set, TS 23 Immunity Development System E2 set
Programmable pulse generation and differential injection
Up to 2.2 kV adjustable | 1.5 ns rise | Adjustable 1.5 kHz repetition | Both polarities | 50 Mbps optical sensor TS 23 included; 24 V / 1 A DC
S2 set, Magnetic Field Probes for E1 S2 set
Transient magnetic-field sensing
Active + passive H-field probes | 3 MSA 02 heads | Optical signal transfer | MSA 02: 1 AAA battery SGZ 21 receiver documented; confirm E2 receiver arrangement
P1 set, Mini Burst Field Generators P1 set
Battery-powered E/H-field injection
3 generators: P11 + P12 + P21 | Approx. 1 mT H-field / 100 kV/m E-field | Single pulse or 5 kHz | 1.5 V AAA 1.5 V AAA supply; single pulse / 5 kHz
P23 set, Mini Burst Field Generator (E) P23 set
Fine-tip capacitive pulse injection
1.2 kV generator voltage | 10 pF coupling | 1.8-10 ns rise | Single pulse or 5 kHz | 1.5 V AAA 1.5 V AAA supply; single pulse / 5 kHz
P11t set, Mini Burst Field Generator (B, Trigger) P11t set
TTL-synchronized magnetic pulse injection
Approx. 1 mT | 5 V TTL SMB trigger | Single pulse to 10 kHz | 3 us trigger delay | 6.5-15 V supply 6.5-15 V supply; power supply included
P12t set, Mini Burst Field Generator (B, Trigger) P12t set
TTL-synchronized clamp-style magnetic coupling
Approx. 1 mT | Coupling-clamp tip | 5 V TTL SMB trigger | Single pulse to 10 kHz | 3 us delay 6.5-15 V supply; power supply included
H2 set, Field Sources H2 set
External-generator E/H-field coupling
6 field sources: 3 H + 3 E | Up to 4.4 kV source supply | External EFT/burst generator required External EFT/burst generator; source limit 4.4 kV
H3 set, Field Sources H3 set
Extended external-generator E/H-field coupling
9 field sources: 4 H + 5 E | High-voltage SMB connections | SHV or Fischer cable choice | External generator required External EFT/burst generator; observe each source rating
PT4 set, EFT Generator Set PT4 set
Potential-free pulse injection plus E/H-field coupling
4.5 kV maximum burst input | 1:1 transformer | 500 V output dielectric strength | H- and E-field sources External EFT/burst generator; transformer maximum 4.5 kV
H4-IC set, EFT/Burst Magnetic Field Source H4-IC set
Small-area magnetic field coupling
Up to 200 mT | 2.54 mm2 resolution | 1.8 mm diameter head | 0-66 MHz | 4 kV maximum burst External EFT/burst generator; 4 kV maximum burst input
H5-IC set, EFT/Burst Magnetic Field Source H5-IC set
Magnetic field coupling with BS 06DU-s
Up to 150 mT | 5 x 6 mm head | 0-560 MHz | 4 kV maximum burst | 15 g source External EFT/burst generator; 4 kV maximum burst input
BD 11, H-Field Burst Detector BD 11
Cable-transient detection and event indication
Adjustable sensitivity | LED + relay + optical outputs | 20 m maximum fiber | 2 AAA batteries | 88 x 85 x 32 mm Two 1.5 V AAA batteries
OTHER PRODUCTS IN THE SAME CATEGORY:
Loading...