DC to 25 kHz | 1 or 2 analog channels | 125 ksps / 12 bit | 0 to 10 V / 0 to 50 V
Langer A100 optical fiber probe sets monitor DUT waveforms during EFT, ESD and RF immunity testing. The included sensor and receiver transmit signals over fiber for display on an oscilloscope. Choose one channel with 1.5 m fiber or two channels with two 6 m fibers. Supported fiber length is 1 to 20 m.
MARKETS:
CATEGORIES:
TEST STANDARDS:
Select Options/Sizes:
Monitor analog voltages inside the device under test while keeping the signal transmission path optical. Select the channel count for the number of signals to observe, then match the sensor range and bandwidth to the waveform.
| Kit | Version | Channels | Included fiber |
|---|---|---|---|
![]() |
A100-1 set | 1 | 1 x 1.5 m |
![]() |
A100-2 set | 2 | 2 x 6 m |
The following sensor and receiver specifications apply to both channel configurations.
| Component | Parameter | Specification |
|---|---|---|
| System | Bandwidth | DC ... 25 kHz |
| System | Sampling rate | 125 ksps |
| System | Resolution | 12 Bit |
| AE 100 Optical Receiver | Bandwidth | DC ... 25 kHz |
| AE 100 Optical Receiver | Optical input: Optical fiber | Ø 2.2 mm |
| AE 100 Optical Receiver | Voltage range - output | 0 V ... 10 V |
| AE 100 Optical Receiver | Supply voltage | 12 V ... 16 V |
| AE 100 Optical Receiver | Current input | ≈ 30 mA |
| AE 100 Optical Receiver | Connector - output | BNC |
| AS 100 Optical Sensor | Bandwidth | DC ... 25 kHz |
| AS 100 Optical Sensor | Sampling rate | 125 ksps |
| AS 100 Optical Sensor | Measuring range | 0 V ... 50 V / 0 V ... 10 V DC (switchable) |
| AS 100 Optical Sensor | Input resistance | 100 k? |
| AS 100 Optical Sensor | Radiated immunity | > 200 V/m |
| AS 100 Optical Sensor | Supply voltage | 3 V ... 16 V |
| AS 100 Optical Sensor | Current input | ≈ 3 mA |
| AS 100 Optical Sensor | Optical fibre length | 1 m ... 20 m |
Specify the sensor when requesting a quote. Check the receiver family and signal range when configuring an alternative to the standard included sensor.
Connect the sensor at the DUT and send its signal through the optical fiber to the receiver. Connect the receiver output to the oscilloscope and supply each module at its specified voltage.

| Kit | Model | Signal | Channels / direction | Bandwidth / data rate | Sampling / function | Supported fiber length | Included fiber |
|---|---|---|---|---|---|---|---|
![]() |
A100-1 set | Analog | 1 | DC to 25 kHz | 125 ksps / 12 bit | 1 to 20 m | 1.5 m |
![]() |
A100-2 set | Analog | 2 | DC to 25 kHz | 125 ksps / 12 bit | 1 to 20 m | 2 x 6 m |
![]() |
A200-1 set | Analog | 1 | DC to 500 kHz | 3 Msps / 12 bit | 1 to 20 m | 1.5 m |
![]() |
A200-2 set | Analog | 2 | DC to 500 kHz | 3 Msps / 12 bit | 1 to 20 m | 2 x 6 m |
![]() |
A300-1 set | Analog | 1 | DC to 5 MHz | 12.5 Msps / 10 bit | 1 to 20 m | 1.5 m |
![]() |
A300-2 set | Analog | 2 | DC to 5 MHz | 12.5 Msps / 10 bit | 1 to 20 m | 2 x 6 m |
![]() |
OSE 150-1 set | Digital | 1 | DC to 50 Mbps | Logic transmission | 1 to 20 m | 1.5 m |
![]() |
OSE 150-2 set | Digital | 2 | DC to 50 Mbps | Logic transmission | 1 to 20 m | 2 x 6 m |
![]() |
OSE 450 set | Digital | 4 | DC to 50 Mbps | Logic transmission | See receiver / sensor configuration | 4 x 6 m |
![]() |
CAN 100 set | CAN | Bidirectional | Up to 1 Mbit/s | High-speed CAN | 10 m maximum; 6 m at 1 Mbit/s | Double 10 m |
![]() |
LIN 100 set | LIN | Bidirectional | LIN bus | Master / slave | Up to 20 m | Double 10 m |
![]() |
OB 100 set | CAN / LIN | Bidirectional | CAN / LIN selectable | 2 Opto-Boxes | 20 m; 6 m at 1 Mbit/s CAN | Double 10 m |