The Multi-CE5 is a compact EMC test unit designed for testing electromagnetic immunity against pulsed and conducted interference. Demonstrating such immunity is generally a requirement for compliance with the European EMC directive, a necessary step leading to the CE mark. In its basic configuration, the Multi-CE5 includes an Electrical Fast Transient Generator (EFTG), a Combination Wave Generator (CWG) and a Coupling-/Decoupling Network (CDN) for single-phase power supply lines.
PARTNER:
MARKETS:
TEST STANDARDS:
Multi CE5
IEC 61000-4-04, IEC 61000-4-05, IEC 61000-4-08, IEC 61000-4-09, IEC 61000-4-11, IEC 60060-2
According to |
|
IEC 61000-4-4 : 2012 |
BURST 5kV |
IEC 61000-4-5 : 2014 |
SURGE 5kV, 2.5kA |
IEC 61000-4-8 : 2010 |
Magnetic field 50/60 Hz |
IEC 61000-4-9 : 2001 |
Magnetic field 8/20 µs |
IEC 61000-4-11 / 29 : 2004 |
Voltage dips/variation |
The Multi-CE5 is a compact EMC test unit designed for testing electromagnetic immunity against pulsed and conducted interference. Demonstrating such immunity is generally a requirement for compliance with the European EMC directive, a necessary step leading to the CE mark.
In its basic configuration, the Multi CE5 includes an Electrical Fast Transient Generator (EFTG-CE5), a Combination Wave Generator (CWG-CE5) and a Coupling-/Decoupling Network (CDN) for single-phase power supply lines.
The Electrical Fast Transient Generator fully compliant to IEC 61000-4-4 delivers fast transient pulses with waveform 5/50 ns and a maximum burst frequency of 1 MHz. It is used for immunity testing of electronic systems and devices. The four standard IEC 61000-4-4 test levels may be easily selected by push button or all parameters may be adjusted individually.
The CWG-CE5 hybrid generator, fully compliant to IEC 61000-4-5 and IEEE 587, is a combined impulse current/ impulse voltage generator that generates a standard impulse with a voltage waveform of 1.2/50µs at high-impedance load (output, RL > 100Ω) and a standard impulse with a current waveform of 8/20µs at short-circuited output. It may be used for surge testing of components and devices, as well as for galvanic coupling of surges to cable shields, shielded enclosures and cabinets.
The built-in capacitive Coupling-/Decoupling Network allows superimposition of the combination wave generator output to the mains voltage of the device under test.
The simulation of voltage dips and voltage variations acc. to IEC 61000-4-11 can be included as an option. Additional accessories allow the testing of immunity against both pulsed and power frequency magnetic fields according to IEC 61000-4-8 and IEC 61000-4-9.
Optionally the Multi-CE5 can include a triggerable power-supply-switch which allows the simulation of the voltage dips as specified in the standard IEC 61000-4-11. The variation of the power supply voltage is controlled by the use of an external motor driven variac. The control of the external power source is included in the mainframe.
An Induction Coil in conjunction with the Combination Wave Generator output is used to simulate pulsed magnetic fields according to IEC 61000-4-9. Combined with the external power source, the Induction Coil can be used to simulate power-frequency magnetic fields according to IEC 61000-4-8.
Additional Coupling-/Decoupling Networks covering three-phase power supply lines, DC supply lines, and signal lines are also available, as well as a Capacitive Coupling Clamp for coupling to shielded interconnection lines.
The Multi-CE5 excels by its compact design, simple handling and precise reproducibility of test impulses. It features a microprocessor controlled user interface and a 7” touch screen unit for ease of use. The microprocessor allows the user to execute either standard test routines or a “user-defined” test sequence. A standard USB port provides the ability to print a summary of the test parameters as well as the results to a USB stick.
The software program CE-REMOTE allows full remote control of the test generator via Ethernet light guide as well as documentation and evaluation of test results, accordingly to the IEC 17025. To record definite impulses, it is equipped with an Impulse Recording Function (IRF).
Moreover all generator functions including the built-in Coupling-/Decoupling Network, maybe computer controlled via the isolated optical interface.
System configuration
Variations |
Description |
Multi-CE5 1 |
including SURGE and BURST |
Multi-CE5 2 |
including SURGE, BURST, and POWER FAIL SWITCH |
EFTG-CE5 |
Stand-alone BURST generator |
CWG-CE5 |
Stand-alone SURGE generator |
PFS-CE-16 |
Stand-alone POWER FAIL SIMULATOR Including a power fail switch and a variable power source |
Typical configurations:
Multi CE5 1 + CDN 5416
for 3-phase testing (picture on the right side)
Multi CE5 2 + VPS 250-16
for testing surge, burst, power fail, voltage dips and variation
It is possible to build all devices in a 19” rack cabinet.
Options |
Multi CE5 |
|
|
Software CE-REMOTE Test, for remote control |
|
With Impulse Recording Function (IRF) |
|
( XP, WIN7, WIN10 ) incl. 5 m fiber optic cable and PC Ethernet interface |
|
|
|
External power source VPS 250-16 |
|
Output voltage, adjustable |
0 - 250 V |
Rated current |
16 A |
Control via interface of Multi CE5 |
|
|
|
Induction Coil HI 200 acc. to IEC 61000-4-8/9: 2010/2001 |
|
Dimensions: W * H * D |
1000*1000*600 mm3 |
Coil factor |
1.5 / m |
|
|
EFTC2012 Coupling Clamp acc. to IEC 61000-4-4:2012 Ed 3.0 |
|
Dimensions: W * H * D |
140 * 180 *1100 mm³ |
Incl. Connection cable, Fischer Koax Connector |
1 m long |
Maximum cable diameter: |
ca. 42mm |
TECHNICAL SPECIFICATIONS |
Multi CE5 |
|
|
Mainframe |
|
Microprocessor controlled touch panel |
7”, capacitive |
Optical Ethernet Interface for remote control of the generator |
optional |
Interface for saving reports |
USB |
External trigger input/ output |
Switch/ 10 V |
Coupling-/decoupling network for power supply lines |
L1, N, PE |
Nominal voltage, nominal current |
250 V, 16 A » / 16 A = |
Coupling impedance (depending on the generator) |
33 nF / 18 µF / 9µF+10W |
Connector for external safety interlock loop |
24 V = |
External red and green warning lamps |
230 V, 60W |
Mains power |
90V - 264V, 50/60 Hz |
Dimensions of desktop case W * H * D |
450*180*500 mm3 |
Weight |
25 kg |
|
|
BURST acc. to IEC 61000-4-4: 2012 |
|
Pulse output voltage, adjustable |
0.2 - 5.0 kV ± 10 % |
Waveform |
5/50 ns |
Source impedance |
50 W |
Polarity, selectable |
pos/neg/alt |
Burst frequency, adjustable |
1.0 kHz - 1.0 MHz |
Burst duration, adjustable |
0,01 ms - 25 ms |
Burst period, adjustable |
10 ms - 1000 ms |
HV output for external coupling devices |
coaxial |
Monitor output for pulse output voltage |
ratio = 100:1 ± 5%, 50 W |
|
|
SURGE acc. to IEC 61000-4-5: 2014 |
|
Test voltage (open circuit condition) |
0.2 - 5.0 kV ± 10 % |
Waveform acc. to IEC 60060 |
1.2 / 50 µs ± 20 % |
Test current (short circuit condition) |
0.1 - 2.5 kA ± 10 % |
Waveform acc. to IEC 60060 |
8 / 20 µs ± 20% |
Polarity of output voltage/current, selectable |
pos/neg/alt |
Maximum stored energy |
120 Joule |
Charging time for max. charging voltage |
< 10 s |
HV output isolated from ground |
HV-OUT, 4mm |
Mains synchronous triggering, phase shifting, digitally selectable |
0 - 359°, step 1° |
Monitor output for pulse output voltage |
ratio = 1000 : 1 ± 5% |
Monitor output for pulse output current |
10 V º 5 kA ± 5% |
|
|
POWER FAIL acc. to IEC 61000-4-11: 2004 |
|
Rated current / Inrush current, max. |
16 A / 500A |
Monitor output for mains voltage and mains current |
built-in |
Display of mains voltage, mains current, and inrush current Interface for control of an external power source |