IPG 1218 / 1012 / 605, Insulation Test, 12kV, 10kV, 6.25kV

High-Voltage Pulse Generators IPG 1218 / 1012 / 605 generate standard impulse voltages with waveform 1.2 / 50 µs acc. to IEC 60060. They are designed for testing impulse dielectric strength of components, insulations, air- and surface flash-over gaps acc. to IEC. The peak value of the test voltage is continuously adjustable from 0.2 - 6.25 / 10 / 12 kV. Positive or negative polarity of the output voltage can be selected. A built-in voltage divider 1000:1 allows monitoring of the impulse output waveform during testing.

PARTNER:

MARKETS:

TEST STANDARDS:

Data Sheet

SPECIFICATIONS

 

IPG  605/ IPG 1012/ IPG 1218

 

  HV - IMPULSE GENERATOR

 

  Surge testing 1.2/50µs

 
  •  0.2-6.25 kV (IPG 605)
  • 0.2-10 kV (IPG 1012)
  • 0.25-12 kV (IPG 1218)

 

 

 

According to

IEC 60664

Testing impulse dielectric strength of components, insulations, air- and surface flash-over gaps

VDE 0110

VDE 0411

VDE 0420

IEC 60335 : 2010

Insulation test of inductors and coils

 

 

High-Voltage Pulse Generators IPG 1218 / 1012 / 605 generate standard impulse voltages with waveform 1.2 / 50 µs acc. to IEC 60060. They are designed for testing impulse dielectric strength of components, insulations, air- and surface flash-over gaps acc. to IEC.

 

The peak value of the test voltage is continuously adjustable from 0.2 - 6.25 / 10 / 12 kV. Positive or negative polarity of the output voltage can be selected. A built-in voltage divider 1000:1 allows monitoring of the impulse output waveform during testing.

 

The generators possess two high-voltage outputs with different source impedance. The HV output terminals are located beyond a dielectric cover with safety interlock. The transparent test cabinet prevents accidental contact with live threatening parts of the test object and allows observation of the test object during testing.

 

 

The generator output possesses a current monitor detecting breakdown or flashover of the test object. The threshold of the current monitor is adjustable.

 

The generator excels by its compact design, simple handling and precise reproducibility of test impulses. It features a microprocessor-controlled user interface and a 7” touch screen unit for ease of use. The microprocessor allows the user to execute either standard test routines or a “user-defined” test sequence. A standard USB port provides the ability to print a summary of the test parameters to a USB stick.

 

The software program IPG-REMOTE allows full remote control of the test generator via Ethernet light guide as well as documentation and evaluation of test results, accordingly to the IEC 17025. To record definite impulses, it is equipped with an Impulse Recording Function (IRF) Moreover, all generator functions may be computer-controlled via the isolated optical interface.

 

 

Options

IPG 605     IPG 1012    IPG 1218

 

PROTECTIVE COVER ON THE EQUIPMENT TOP

With safety interlock switch, connected to the safety interlock loop,

red and green warning lamps installed acc. VDE 0104

See figure

Type PA 503, Dimensions W * H * D

400 * 140 * 300 mm³

Type PA 505, Dimensions W * H * D

400 * 250 * 400 mm³

 

Software IPG-REMOTE, for remote control

With Impulse Recording Function (IRF)

( XP, WIN7, WIN10 ) incl. 5m long light guide and PC Ethernet interface

 

Version without protective cover, current shunt Rm = 1W,

BNC for measuring on the back

 

 

 

TECHNICAL SPECIFICATIONS

IPG 605

IPG 1012

IPG1218

 

 

Mainframe

 

Microprocessor controlled touch panel

7”, capacitive

Optical Ethernet Interface

for remote control of the generator

Optional

Interface for saving reports

USB

External Trigger input/ output

Switch/ 10V

Connector for external safety interlock loop

24 V=

External red and green warning lamps

24 V=, 40 mA

Mains power

90V – 264V / 50/60 Hz

Dimensions of desktop case W * H * D

450*180*500 mm3

Weight

18kg

 

 

Generator section

 

Peak value of impulse output voltage, adjustable, ± 5 %

0.2-6.25 kV

0.2-10 kV

0.25-12 kV

Waveform of impulse output voltage, acc. to VD 0433, IEC 60060

1.2/50 µs ± 30 % / 20 %

Max. stored energy

5 Joule

12 Joule

18 Joule

Energy storage capacitor                           Cs

0.25 µF

Resistor in series to the output HV1          Rs1

500W

Resistor in series to the output HV2          Rs2

Standard 50W (opt. 40 oder 200W)

Output polarity, selectable

pos / neg / alt

 

 

Trigger :

 

a) manual

Push button

b) external Trigger input

Switch

c) internal, automatic, adjustable via test procedure

1 - 1000 pulses

Repetition time, selectable

1-1000 s

3 -1000 s

3 -1000 s

 

 

CURRENT SENSE

built-in

Threshold value, selectable

1-830µAs

1-1200 µAs

1-1500 µAs

Impulse voltage divider, built-in

1000:1 ± 2 %

Mains synchronous triggering, phase shifting, digitally selectable

0 - 359°, step 1°

HV output, HV-OUT

HV connector

 

 

Accessories: power cable, turn key, instruction manual

 