HTS 20-10, 20 kV, DC dielectric strength

High-Voltage DC Test Sets HTS 20-10 are designed for testing dc dielectric strength of components, insulations, air- and surface flash-over gaps. Also for testing SPDs according to IEC 61643-1. The test voltage can be adjusted from 0.2 kV up to 20 kV. The test voltage rises like a ramp from zero to the selected test voltage. The rate of rise can be adjusted from 100 V/sec up to 10000 V/sec. Then the test voltage remains at the selected value for the test time, which can be adjusted from 1 – 1000 sec. Figure: incl. option PA503

PARTNER:

MARKETS:

TEST STANDARDS:

Data Sheet

SPECIFICATIONS

 

HTS 20-10

 

 

High voltage test of components

 

 

 
  •  High Voltage AC Test Set
 
  • Voltage: 0.2 - 30 kV
 
  • Current: 1 - 20 mA
 
  • Rise: 0.2 - 30 kV/sec
 
  • Testtime: 1 - 100 sec

 

 

Limit current,  adjustable, Error message on over current

Current control in mode “current”: 0.1-10mA

 

 

 

High-Voltage DC Test Sets HTS 20-10 are designed for testing dc dielectric strength of components, insulations, air- and surface flash-over gaps. Also for testing SPDs according to IEC 61643-1. The test voltage can be adjusted from 0.2 kV up to 20 kV.

 

The test voltage rises like a ramp from zero to the selected test voltage. The rate of rise can be adjusted from 100 V/sec up to 10000 V/sec. Then the test voltage remains at the selected value for the test time, which can be adjusted from 1 – 1000 sec. The actual test

voltage is displayed. The test can be repeated several times. If the output current exceeds the limit value, which can be adjusted, the high-voltage power supply is switched off and an error message is generated: sound signal and display. If an over-current occurs during rise of voltage, the spark-over The voltage is displayed.

 

In "current" mode, the test voltage increases in a ramp to the set current value.

Then the generator regulates the current value and measures the voltage at the test object. The adjustment range of the current is from 0.1 - 10 mA.

 

 The HV output terminals are located beyond a dielectric safety test cover with safety interlock. The transparent test cabinet prevents accidental contact with live parts of the test object and allows observation of the test object during testing.

 

It features a microprocessor-controlled user interface and a 7” touch screen unit for ease of use. The microprocessor allows the user to execute either standard test routines or a “user-defined” test sequence. A standard USB port provides the ability to print a summary of the test parameters to a USB stick.

 

The software program HTS-REMOTE allows full remote control of the test generator via Ethernet light guide as well as documentation and evaluation of test results, accordingly to the IEC 17025. To record definite impulses, it is equipped with an Impulse Recording Function (IRF) Moreover all generator functions may be computer-controlled via the isolated optical interface.

 

The generator excels by its compact design, simple handling and precise reproducibility of test impulses.

 

 

Options

HTS 20-10

 

PROTECTIVE COVER ON THE EQUIPMENT TOP

With safety interlock switch, connected to the safety interlock loop,

red and green warning lamps installed acc. VDE 0104

See figure

Type PA 503, Dimensions W * H * D

400 * 140 * 300 mm³

Type PA 505, Dimensions W * H * D

400 * 250 * 400 mm³

 

Software HTS-REMOTE, for remote control

With Impulse Recording Function (IRF)

( XP, WIN7, WIN10 ) incl. 5m long light guide and PC Ethernet interface

 

TECHNICAL SPECIFICATIONS

HTS 20-10

 

 

Mainframe:

 

Microprocessor controlled touch panel

7”, capacitive

Optical Ethernet Interface

for remote control of the generator

Optional

Interface for saving reports

USB

External Trigger input/ output

Switch/ 10V

Connector for external safety interlock loop

(not with Opt. 1)

24 V=

External red and green warning lamps

230V / 60 W

Mains power

90V – 264V / 50/60 Hz

Dimensions of desktop case W * H * D

450*180*500 mm3

Weight

25kg

 

 

High-Voltage Test Set Mode turn-off:

 

Test voltage

0.2 - 20 kV(± 5%)

Measurement of test- /spark-over voltage/current, accuracy

± 5%

Voltage rate of rise, adjustable

100 V/sec-10 kV/sec

Test time, adjustable

1 - 1000 sec

Limit current, adjustable

0.1-10mA

Number of repetitions, adjustable

1 - 1000

High-Voltage Test Set Mode current:

 

Regulated current

0.1 - 10 mA(± 5%)

Measurement of test- /spark-over voltage/current, accuracy

± 5%

Voltage rate of rise, adjustable

100 V/sec-10 kV/sec

Test time, adjustable

1 - 1000 sec

Number of repetitions, adjustable

1 - 1000

 

 

HV output terminals, on the top of the equipment

2 banana plugs, 4 mmØ

 

OTHER PRODUCTS IN THE SAME CATEGORY: