IGUU 2918, Calibration-Pulse Generator for CISPR 16 Receivers

  • Calibration-Pulse Generator acc. CISPR 16
  • For Band A, B, C, D (9 kHz-1000 MHz)
  • To calibrate the pulse response of the EMI receiver
  • Pulse repetition frequency main generator 0.1 - 200 Hz (aux. generator up to 1 MHz) with IEEE-488 Interface.

MARKETS:

TEST STANDARDS:

Manual

Differences between IGUU 2916 & 2918

KU 9618 Coax switch Option

Software Option

SPECIFICATIONS

 

IGUU 2918, Calibration-Pulse Generator for CISPR 16 Receivers


  1. The enhanced auxiliary generators have an extended pulse frequency range from 0.1Hz to 1MHz (Band C/D). Now it is possible to test all previous detectors and additionally also True RMS detectors.
  2. Rotary encoders and matrix - displays are used for pulse frequency and amplitude (main generator only) tuning and reading. The new keypad allows a direct entry of the desired frequency and amplitude. Pulse frequency is derived from a DDS-IC.
  3. The optional Coaxial Switching Unit KU 9618 connects one of the four outputs of the IGUU 2918 to the e. u. t., controlled by the IGUU 2918. This is very convenient and time-saving because there is no need to change coaxial connections.
  4. The GPIB - Interface serves for complete PC-control of all generator functions. Automatic testing is possible by using the interface and coaxial switching unit KU 9618.

 

 

  1. Main Generators
  • Pulse frequency range: 0.1Hz - 200Hz in 0.1Hz steps
  • Pulse frequency accuracy: 2.3·10-3Hz Pulse shape: Almost rectangular Pulse duration: Band A: 270ns,
  • Band B: 6ns Band C/D: 0,3ns
  • Pulse polarity: positive/negative (switchable)
  • Pulse level: (CISPR weighting Q.P.): Band A / B 10 - 70dB, in 1dB - steps, Band C / D 0 - 60dB, in 1dB - steps
  • Output Voltage acc. 50Ω: 80 V - 110 V
  • Pulse Area at 60dB attenuator setting: Corresponds to CISPR 16 Standard equivalent to 1 mVrms across 50 Ω with standardized Q.P. - Detector. 
E*t Band A 9kHz - 150kHz 6.75 µVs
E*t Band B 150kHz - 30MHz 0.158 µVs
E*t Band C / D 30MHz - 1000MHz 0.022 µVs
  • Frequency flatness:
    • + / - 0dB to - 0.5dB at upper frequency limit for Band A / B
    • Band C: typ. + / - 0,5dB 30 - 100MHz (100MHz is reference frequency for level calibration).
    • Level increase between 200MHz and 800MHz + 0,5 to + 1dB, 800 - 1000MHz decrease to + / - 0dB to - 1dB.
  • LEVEL ACCURACY:
    • Band A / B: + / - 0,5dB at 50kHz, 1MHz
    • Band C / D: + / - 0,8dB at 100MHz.
  1. Auxiliary generators
  • Pulse frequency ranges:
    • Band A: 0.1Hz – 2kHz
    • Band B: 0.1Hz – 50kHz
    • Band C/D: 0.1Hz – 1000kHz
      • 1Hz – 999.9Hz, 0.1Hz steps
      • 1kHz – 1000kHz, 0.1kHz steps
    • Pulse frequency accuracy: 2.3x10^-3Hz
    • Pulse duration (values for orientation) :
      • Band A: 260ns,
      • Band B: 6.5ns,
      • Band C / D: 500ps
    • Pulse level: ((CISPR weighting Q.P.):
      • Band A / B 60dB fixed level
      • Band C / D 40dB fixed level
    • Frequency flatness:Band A / B same as 1. Main Generators
      • Band C (D): typ. + / - 0,75dB 30 – 1000MHz
  1. Dimensions: W x H x D
  • 447mm x 180mm x 460mm
  1. Weight: 12kg

 Options:

Opt. RecTest Softw.

Receiver Test Software for IGUU 2916 or 2918 signal generator and EMI-receiver to perform automatic calibration of an EMI receiver acc. to CISPR 16-1-1.

 

Opt. KU 9618

Coaxial Switching Unit for automatic performance tests with IGUU 2918, N-Connectors female.

 

 

OTHER PRODUCTS IN THE SAME CATEGORY: