The high-voltage measurement instruments HVM 2015 consist of a measuring amplifier for universal use with a digital touch display, a built-in high-voltage divider for direct measurement up to 10 kV / optional 20 kV and an external input for an external high-voltage divider, HVT***RCR, measurement, and display of voltages up to 300 kV. The HVM 2015 is designed for measuring ac- and dc- voltages and allows in addition measurement of the peak value of lightning and switching pulse voltages.
High-voltage measurement equipment
HVM 2015
The high-voltage measurement instruments HVM 2015 consist of a measuring amplifier for universal use with a digital touch display, a built-in high-voltage divider for direct measurement up to 10 kV / optional 20 KV and an external input for an external high- voltage divider, HVT***RCR, measurement, and display of voltages up to 300 kV.
The HVM 2015 is designed for measuring ac- and dc- voltages and allows in addition measurement of the peak value of lightning and switching pulse voltages.
The device features a microprocessor-controlled user interface and a 5” touch screen unit for ease of use. It allows the user to set parameters and shows the results on the display.
A standard USB port provides the ability to print a summary of the test parameters to a USB stick.
All HVT***RCR voltage dividers are predefined in the setup for easy use, plugging and measuring. Furthermore, other extern dividers can be connected by using the freestyle setup.
The optional software program REMOTE allows full remote control of the test HVM2015 via Ethernet light guide as well as documentation and evaluation of test results, accordingly to IEC 17025. It is equipped with an Impulse Recording Function (IRF) for recording definite impulses.
Optionally the unit can be delivered with a DKD calibration certificate.
|
Technical specification: |
HVM 2015 |
Microprocessor controlled touch screen |
5”, 800X480, 24 bit |
Optical-Interface for remote control of the generator |
optional |
Interface for saving reports |
USB |
External trigger input /output |
10 V at 1 kW |
Mains power |
230 V/50 Hz / 10 VA |
Dimensions of desk top case W * H * D |
450 * 180 * 300 mm3 |
Weight |
5kg |
|
|
Direct high-voltage-input |
10 kV= / 7kV» |
External input, Ri = 1000 kW |
100Vº100 % |
Calibration possibility of the preamplifier |
± 2 % |
Analogue output, Ri = 5 kW |
5 V º 100 % 1 mA º 100 % |
|
|
Setting Parameter / Evaluation input voltage |
|
DC |
Average |
AC |
+ / - Peak/Ö2 |
Impulse |
+ / - Peak |
Impulse |
Reset |
Measurement input |
Intern / Extern |
Time interval for storing the measurement value |
1s – 3600s |
|
|
Resolution AD-converter |
12 Bit / 4096 |
Digital display |
Autorange |
Precision AC/DC |
1.0 % |
Precision impuls |
2.5 % |
Bandwidth |
0 -10 kHz |
Environmental temperature |
17 °C - 23 °C |
|
|
Options |
|
Direct high-voltage-input, optional |
20 kV= /14kV» |
PC Remote Control SW with light guide, Ethernet Interface |
|
|
|
|