The High-Voltage Test-Set HVTS 5010 is designed for high-voltage testing of the electrical insulation of electrical equipment and electrical systems. The test voltage is adjustable up to 50 kV, 1 - 10 mA. The High-Voltage Test-Set HVTS 5010 mainly consists of a high-voltage transformer and a control unit. A motor-driven variac generates the primary voltage of the high-voltage transformer. The output voltage can be selected from 2.0 to 50 kV with a voltage rise adjustable from 3 kV/sec to 50 kV/sec. The maximum output current is 10 mA. The output voltage is measured by a built-in capacitive voltage divider.
High-Voltage AC-Test-Set HVTS 50-10
The High-Voltage Test-Set HVTS 5010 is designed for high-voltage testing of the electrical insulation of electrical equipment and electrical systems. The test voltage is adjustable up to 50 kV, 1 - 10 mA. The High-Voltage Test-Set HVTS 5010 mainly consists of a high-voltage transformer and a control unit. A motor-driven variac generates the primary voltage of the high-voltage transformer. The output voltage can be selected from 2.0 to 50 kV with a voltage rise adjustable from 3 kV/sec to 50 kV/sec. The maximum output current is 10 mA. The output voltage is measured by a built-in capacitive voltage divider.
The HVTS 5010 features a microprocessor-controlled user interface and a 5” touch screen unit for ease of use. The microprocessor allows the user to execute either standard test routines or a “user defined” test sequence. A standard USB port provides the ability to print a summary of the test parameters to a USB stick.
Technical specification: |
HVTS 50-10 |
Mainframe control unit: |
|
Microprocessor controlled touch panel |
5”, 800X480, 24 bit |
Optical Ethernet Interface for remote control of the generator |
optional |
Interface for saving reports |
USB |
Connector for external safety interlock loop |
24 V = |
external red and green warning lamps acc. to VDE 0104 |
230 V, 60W |
|
|
HV unit: |
|
Test voltage, adjustable |
2 - 50 kV» |
Accuracy of the test voltage |
< 2% |
Rise of sinus voltage ramp, adjustable |
3 - 50 kV/sec |
Output current |
1.0 - 10.0 mA |
Accuracy of the test current |
< 2% |
Short circuit current |
> 20.0 mA |
High-voltage output, one terminal connected to GND |
hv bushing |
Connection for ground rod |
M6 |
Connector for external safety interlock loop |
24 V = |
and external red and green warning lamps acc. to VDE 0104 |
230 V, 60W |
Mains power |
230 V, 50/60 Hz |
Dimensions W * H * D |
553*800*1250 mm3 |
Weight |
80 kg |
Option: |
|
Test chamber on top, build in 19” rack, with security glass door, safety interlock protects the high-voltage output terminals. Upon opening of the door or switching-off of the generator, the HV output switch off. Test space ca. W*H*D |
470*530*490 mm³ |