HHS 5210-100 2.5 - DC - 10 kHz, Square Helmholtz Coils

Square Helmholtz Coils

Precisely defined magnetic fields. DC - 10 kHz.

The Schwarzbeck HHS 5210-100-2.5 generates defined magnetic fields from DC to 10 kHz. Its 100-turn coils provide a calculable relationship between drive current and magnetic field strength.

MARKETS:

TEST STANDARDS:

NFCN 9732-120 - Compensation Network

Data Sheet

SPECIFICATIONS

 DC - 10 kHz

Usable frequency range
12 AContinuous nominal coil current
100 turnsPer coil
Defined by geometry

A calculable field.
A traceable measurement.

Magnetic field strength is proportional to coil current. Coil geometry, the number of turns and the applied current determine the resulting field, allowing the field measurement to be traced to a current measurement or the voltage drop across a known resistor.

The field depends on coil spacing and the position within the test volume. Use the reference spacings in the specifications and the field distribution data in the datasheet when selecting your setup.

Plan your test setup

Match the field to your application.

Continuous operation

The nominal coil current is 12 A continuous. The specified nominal magnetic field is 1746 A/m continuous. Field values apply at the stated reference spacing.

Short-duration operation

The maximum coil current is 20 A (5 min.). Observe the stated time limit and provide adequate ventilation. The frequency range alone does not imply that maximum current or field is available at every frequency.

Technical details

Specifications

Product specifications with reference conditions. See the linked manufacturer datasheet for complete field distributions and operating guidance.

View the complete datasheet →
Number of turns 100
Maximum Coil Current 20 A (5 min.)
Nominal Coil Current 12 A continuous
Inner side length of the coil 1.0 m
Electrical side length of the coil 1.04 m
Maximum Coil Separation 0.70 m
Usable Frequency Range DC - 10 kHz
Inductance (per Coil) 31.3 mH
Resonance Frequency (Pair of Coils) >25 kHz
Weight 56 kg
Terminals 4 mm female
Coil Separation acc. IEC 1000-4-8 0.40 m
Max. Magn. Field Strength 2900 A/m (5 min.)
Nominal Magn. Field Strength 1746 A/m continuous
Magn. Field Strength, 1 A Coil Current 145.54 A/m (Coil Dist. 0.4 m)
Coil Separation for best uniformity along an axes through the coil center 0.566 m
Magn. Field Strength, 1 A Coil Current 124.66 A/m (Coil Dist. 0.566 m)
Nominal/maximum field reference spacing 0.40 m
Mechanical Dimensions 1.15 m × 1.20 m × 0.75 m

Where specified, coil distances refer to the coil centers. Continuous and short-duration limits are listed separately. Standards references retain the designations used in the source documents.

NFCN 9732-120 - Compensation Network

The compensation network NFCN 9732-120 consists of a capacitor and a fuse in series. It is used for series compensation of the HHS 5210-100 at electrical mains frequencies of 50 Hz to 60 Hz.

 

The compensation capacitor reduces the total impedance of the series circuitry consisting of the Helmholtz coils HHS 5210-100 and the NFCN 9732-120 at the operating frequencies of 50 to 60 Hz. It allows continuous currents of up to 8 A rms at generator voltages of less than 70 Vrms ...80 Vrms 

Installation and operating safety

Read the supplied datasheet before use. Switch off the complete system before making or removing connections and do not touch terminals during operation. Observe all voltage, current and time limits. High currents heat the coils; maintain adequate air circulation. Keep unintended magnetic field sources and magnetic materials away from the test volume.

North American sales

Discuss your test requirements.

Contact Absolute EMC LLC for the HHS 5210-100-2.5 in the United States, Mexico and Canada.

OTHER PRODUCTS IN THE SAME CATEGORY: