Coil Set 4-39_150k-26M EN 61000-4-39, 150 kHz - 26 MHz, Tx & Rx Set

High-Frequency Transmit Coil (CAB4-3910)

  • Frequency: 150 kHz to 26 MHz
  • Main Diameter: 100 mm
  • Number of Turns: 3
  • Wire Diameter: 1mm
  • Connection type: ‘N’ type Coaxial

High-Frequency Sense Coil (CABLSHF)

 

  • Frequency: 150 kHz to 26 MHz
  • Diameter to center of the coil: 41 mm
  • Number of turns: 1
  • Wire type: 1mm single core
  • Winding Type: Electrostatically shielded

Manual

Data Sheet

SPECIFICATIONS

Coil Set 4-39_150k-26M For testing IEC 61000-4-39 and IEC 60601-1-2, 150 kHz - 26 MHz

Introduction
The Conformity Assessment Business EN61000-4-39:2017 test set consists of two transmit coils, and two complimentary electrostatically shielded sensor coils designed as per section 6.1 of the standard. These cover the frequency ranges 9 kHz - 150 kHz and 150 kHz - 26 MHz.
For the high frequency transmit coil, an additional coupling network is supplied to enable the 7.5 A/m test level of EN60601-1-2 to be met. This should be connected in-line at the transmit probe head end.

High Frequency Transmit Coil (CAB4-3910)

Specification  
Frequency: 150 kHz - 26 MHz
Main Diameter: 100 mm
Number of turns:  3
Wire Diameter: 1 mm
Connector: N Type Coax

 

High Frequency Sense Coil (CABLSHF)

Specification  
Frequency: 150 kHz - 26 MHz
Diameter to center of the coil:  41 mm
Number of turns:  1
Wire type:  1mm Single Core
Winding Type: Electrostatically shielded
Connector: BNC


 For testing 13.56MHz 7.5A/m levels of IEC 60601-1-2 requirments a matching network is required. Coupling Network 13.56M
 

OTHER PRODUCTS IN THE SAME CATEGORY:
Loading...