Coil Set 4-39_9k-150k EN 61000-4-39 & RS101 9kHz - 150kHz, Tx & Rx Set

Low-Frequency Transmit Coil (CAB4-3912)

  • Frequency: 9 kHz to 150 kHz
  • Main Diameter: 120 mm
  • Maximum Current: 15 A
  • Number of turns: 20

Low-Frequency Sense Coil (CABLSLF) 

  • Frequency: 9 kHz to 150 kHz
  • Diameter to center of the coil: 40 mm
  • Number of turns: 51
  • Wire type: 7/41 LITZ
  • Winding Type: Electrostatically shielded

Manual

Data Sheet

SPECIFICATIONS

Coil Set 4-39_9k-150k For testing RS101, IEC 61000-4-39 and IEC 60601-1-2, 9kHz - 150kHz

Introduction
The Conformity Assessment Business EN61000-4-39:2017 test set consists of two transmit coils and two complimentary electrostatically shielded sensor coils designed as per section 6.1 of the standard. These cover the frequency ranges 9 kHz - 150 kHz and 150 kHz—26 MHz. The Design for the COIL SET 4-39_9k-150k is the same as referenced in MIL-STD-461 RS101 testing. 
An additional coupling network is supplied for the high-frequency transmit coil to enable the 7.5 A/m test level of EN60601-1-2 to be met. This should be connected in-line at the transmit probe head end.

Low-Frequency Transmit Coil (CAB4-3912)

Specification  
Frequency: 9 kHz to 150 kHz
Main Diameter: 120 mm
Maximum Current: 15 A
Number of turns:  20
Wire Diameter: 2.1 mm
Connector: Banana 4mm

Matching network for 134 kHz CAB134k

 

Low-Frequency Sense Coil (CABLSLF)

Specification  
Frequency: 9 kHz to 150 kHz
Diameter to center of the coil:  40 mm
Number of turns:  51
Wire type:  7/41 LITZ
Winding Type: Electrostatically shielded
Connector: BNC

Test standards:

  • MIL-STD-461D/E/F RS 101, Magnetic immunity testing. 
  • IEC 61000-4-39, Magnetic immunity to proximity items
  • IEC 60601-1-2, Medical EMC Requirments, proximity testing for RFID referencing IEC 61000-3-39

 