HHS 5202-9 - DC - 2.5 MHz, Circular Helmholtz Coils

  • Circular Helmholtz Coils
  • DC - 2.5 MHz
  • Diam. 232mm
  • turns 9

MARKETS:

TEST STANDARDS:

Data Sheet

SPECIFICATIONS

 

HHS 5202-9 Circular Helmholtz Coils   

Helmholtz-Coils are especially designed to generate precisely defined magnetic fields from DC to the upper end of the audio frequency range and beyond. The generated fields are in strong linear relation to the coil current. The field strength can be calculated exactly by analytical (or numerical) methods, based on the coils' geometry, the number of turns, and the coil current. Therefore the HHS 5202-9 is ideally suited for the calibration of magnetic field probes or sensors. Due to the high temperature proof Teflon-coated wire packet, it is possible to generate magnetic fields up to approx. 3000 A/m (for a short time). Typical applications are magnetic immunity testing according to automotive standards or MIL-461F. When generating magnetic fields with Helmholtz coils the coil current is directly proportional to the magnetic field strength.

The calibration of the magnetic field is finally traceable to a current measurement (or to a voltage drop at a known resistor).

The Helmholtz Coil itself is usually considered as a primary standard due to the easily calculable relation between current and field strength. If this relation should be controlled, a loop sensor or monitoring loop can be used to determine the actual field strength.

The Helmholtz-Coils should be installed on a desk in a sufficiently large separation from sources of unintentional magnetic fields, e.g. transformers in power supplies, conductors carrying high currents, computer monitors, loudspeakers, cathode ray tubes (CRT), and more.... All kinds of magnetic material (e.g. steel, Nickel, Cobalt) should be removed from the near surrounding of the coil. The wires which are used to connect the current source with the Helmholtz- Coil should be twisted to avoid an unwanted injection of magnetic flux. The coil terminals are assigned with the characters A, B, C, and D. The generator (current source, audio-amplifier...) is connected to the terminals A and C, the terminals B and D are connected with the short cable supplied with the coil.

Specifications:       

  Number of turns: 9
  Maximum Coil Current: 55 A (5 min.)
  Nominal Coil Current: 33 A continuous
  Coil Spacings (centered): 116 mm / 132 mm
       
  The following data refers to a coil separation of 116 mm
       
  Max. Magn. Field Strength: 3053 A/m (5 min.)
  Nominal Magn. Field Strength: 1832 A/m continuous
  Magn. Field Strength, 1 A Coil Current:

55.52 A/m
154.89 dBµA/m

  Coil Diameter: 232 mm
  Mechanical Dimensions: 0.20 m x 0.25 m x 0.39 m
  Max. cubical shaped DuT: 11.3 cm x 11.3 cm x 11.3 cm
  Terminals: 4 mm female jacks
7 mm screw terminals
  Inductance (Single Coil): 36 µH
  Inductance (Coil Pair):

80 µH

  Usable Frequency Range: DC - 2.5 MHz
  Resonance Frequency: > 3.5 MHz
 

Weight:

4.9 kg

 

  

 

 

 