TEM-6000 STRIPLINE, 6 GHz IC Testing

TEM-6000 STRIPLINE Mod.2020 is a 6 GHz IC Stripline TEM Cell to generate EM fields for testing sample IC devices. It meets the requirements of IEC 61967-2/8, IEC 62132-8 and SAE 1752-3

SPECIFICATIONS

TEM-3000, TEM-6000 STRIPLINE Ic Testing

TEM3000 and TEM6000 Striplines are 3GHz and  6 GHz IC Stripline TEM Cell to generate EM fields for testing sample IC devices. They meet the requirements of IEC 61967-2/8, IEC 62132-8, and SAE 1752-3.

MARKETS: Automotive, CE Mark (Industrial & Residential), Components, Industrial, INFORMATION TECHNOLOGY, Medical Electronics, Military, Power Generation, RF, Wireless/Telecom,


TEST Standards
  • IEC 62132-8: 2012 Integrated circuits – Measurement of electromagnetic immunity, 150 kHz up to 3 GHz – Part 8: Measurement of radiated immunity – IC stripline method
  • IEC 61967-8: 2011 Integrated circuits – Measurement of electromagnetic emissions, 150 kHz up to 3 GHz – Part 8: Measurement of radiated emissions – IC stripline method
  • IEC 61967-8: 2011 Integrated circuits – Measurement of electromagnetic emissions, 150 kHz up to 3 GHz – Part 8: Measurement of radiated emissions – IC stripline method
  • SAE 1752-3 Measurement of Radiated Emissions from Integrated Circuits — TEM/Wideband TEM (GTEM) Cell Method; TEM Cell (150 kHz to 1 GHz), Wideband TEM Cell (150 kHz to 8 GHz)
  • Electromagnetic Radiation test of IC
  • Electromagnetic Immunity test of IC
 
Description
TEM-6000 STRIPLINE Mod.2020 is a DC-6 GHz IC Stripline TEM Cell for testing small devices such as IC’s, wireless communication modules, etc. An external test signal applied through the input port of the TEM-6000 STRIPLINE Mod.2020 generates a consistent and predictable TEM test field inside the cell. The radiation field from a device transmitting in the Cell can also be detected through the port using a test receiver. Requires 50Ω load for termination. The high-quality compact design is optimized for measurements beyond the standard TEM Cell frequency range to 6 GHz. The E-H field inside the test volume is proportional to the input voltage and inversely proportional to the cell height. Since this apparatus is very broadband, it has many applications in the area of EMI, EMS, receiver sensitivity tests, etc.

Features:
  • Up to 6 GHz bandwidth (Beyond normal TEM Cell bandwidth of 1 GHz)
  • Can handle up to 5 kV high voltage pulse for Transient Field Injection Test
Technical Specifications  
Frequency range DC – 3GHz, DC- 6 GHz
TEM Cell Impedance 50Ω ± 5% nominal
VSWR DC- 3 GHz  <1.2 | 3 – 5 GHz <1.5 |5 – 6 GHz <1.9
Insertion Loss (S21) Less than 0,2dB
Return Loss (S11 & S22) DC- 3 GHz  >20 dB | 3 – 5 GHz >14 dB | 5 – 6 GHz >10 dB
RF Connectors N-Type up to 3GHz, and SMA up to 6GHz
Maximum Input Power 500 Watts up 3GHz, 200W up to 6GHz
Maximum Input Voltage 3 kV @DC,  5kV @Pulse
Effective Cell Height 30 mm up to 3GHz and 12.5 mm up to 6GHz
DUT Port Dimensions 80(W)x80(D) and 50 (W) x 50 (D) mm
Recommended MAX DUT Dimensions 30 (W) x 30(W) x 3(H) mm
TEM Cell Dimensions 170 (W) x 120 (D) x 70 (H) mm;6.5 (W) x 4.5 (D) x 1 in.
Weight Approx 2 kg;4 lbs.

 

 

OTHER PRODUCTS IN THE SAME CATEGORY: