50 µH 16 A LISN with Switchable RF Protection
MARKETS:
CATEGORIES:
TEST STANDARDS:
TBL5016-2B Single-phase LISN with independent line and neutral BNC outputs for conducted-emissions work and simultaneous connection to a LISN Mate.
| Measurement range | 9 kHz–30 MHz |
|---|---|
| Network impedance | 50 Ω ? (50 µH + 5 Ω) |
| Maximum line voltage | up to 250 V AC at 50/60 Hz or DC |
| Current rating | 16 A at 23 °C subject to socket rating |
| RF outputs | two 50 Ω BNC outputs |
| Power inlet | IEC 60320 C19 power inlet |
| Artificial hand | 220 pF + 511 Ω artificial hand |
| Protective earth | switchable PE |
| Operating environment | +5 to +40 °C, 5–80% RH |
CISPR 16-1-2 conducted-emissions measurements.
Specify EU, US, AU or UK EUT socket. US NEMA 5-15 and AU options are rated 15 A; UK is rated 13 A. The manual lists a mains cord, coaxial measurement cable and BNC termination with the unit. Select external RF input protection for the receiver as needed.
TBL5016-2C Single-phase LISN with a line/neutral selector and switchable limiter, attenuator and high-pass filter in the measurement path.