Continuous operation
The nominal coil current is 10 A continuous. The specified nominal magnetic field is 125 A/m continuous. Field values apply at the stated reference spacing.
Precisely defined magnetic fields. DC - 150 kHz.
The Schwarzbeck HHS 5212-10 generates defined magnetic fields from DC to 150 kHz. Its 10-turn coils provide a calculable relationship between drive current and magnetic field strength.
Magnetic field strength is proportional to coil current. Coil geometry, the number of turns and the applied current determine the resulting field, allowing the field measurement to be traced to a current measurement or the voltage drop across a known resistor.
The field depends on coil spacing and the position within the test volume. Use the reference spacings in the specifications and the field distribution data in the datasheet when selecting your setup.
The nominal coil current is 10 A continuous. The specified nominal magnetic field is 125 A/m continuous. Field values apply at the stated reference spacing.
The maximum coil current is 20 A (5 min.). Observe the stated time limit and provide adequate ventilation. The frequency range alone does not imply that maximum current or field is available at every frequency.
Product specifications with reference conditions. See the linked manufacturer datasheet for complete field distributions and operating guidance.
View the complete datasheet →| Number of turns | 10 |
|---|---|
| Maximum Coil Current | 20 A (5 min.) |
| Nominal Coil Current | 10 A continuous |
| Maximum Coil Separation | 0.70 m |
| Usable Frequency Range | DC - 150 kHz |
| Inductance (per Coil) | 0.45 mH |
| Resonance Frequency (Pair of Coils) | >500 kHz |
| Weight | 12 kg |
| Terminals | 4 mm female |
| Coil Separation acc. IEC 1000-4-8 | 0.48 m |
| Coil Separation for best uniformity | 0.68 m |
| Max. Magn. Field Strength | 250 A/m (5 min.) |
| Nominal Magn. Field Strength | 125 A/m continuous |
| Magn. Field Strength, 1 A Coil Current | 12.45 A/m (Coil Dist. 0.48 m) |
| Mechanical dimensions | 1.2 m × 1.2 m × 0.74 m |
Where specified, coil distances refer to the coil centers. Continuous and short-duration limits are listed separately. Standards references retain the designations used in the source documents.
Read the supplied datasheet before use. Switch off the complete system before making or removing connections and do not touch terminals during operation. Observe all voltage, current and time limits. High currents heat the coils; maintain adequate air circulation. Keep unintended magnetic field sources and magnetic materials away from the test volume.
Contact Absolute EMC LLC for the HHS 5212-10 in the United States, Mexico and Canada.