HHS 5206-16 - DC - 500 kHz, Circular Helmholtz Coils

Circular Helmholtz Coils

Precisely defined magnetic fields. DC - 500 kHz.

The Schwarzbeck HHS 5206-16 generates defined magnetic fields from DC to 500 kHz. Its 16-turn coils provide a calculable relationship between drive current and magnetic field strength.

MARKETS:

TEST STANDARDS:

Data Sheet

Security Note

SPECIFICATIONS
DC - 500 kHzUsable frequency range
33 AContinuous nominal coil current
16 turnsPer coil
Defined by geometry

A calculable field.
A traceable measurement.

Magnetic field strength is proportional to coil current. Coil geometry, the number of turns and the applied current determine the resulting field, allowing the field measurement to be traced to a current measurement or the voltage drop across a known resistor.

The field depends on coil spacing and the position within the test volume. Use the reference spacings in the specifications and the field distribution data in the datasheet when selecting your setup.

Plan your test setup

Match the field to your application.

Continuous operation

The nominal coil current is 33 A continuous. The specified nominal magnetic field is 1260 A/m continuous. Field values apply at the stated reference spacing.

Short-duration operation

The maximum coil current is 55 A (5 min.). Observe the stated time limit and provide adequate ventilation. The frequency range alone does not imply that maximum current or field is available at every frequency.

Technical details

Specifications

Product specifications with reference conditions. See the linked manufacturer datasheet for complete field distributions and operating guidance.

View the complete datasheet →
Number of turns 16
Maximum Coil Current 55 A (5 min.)
Nominal Coil Current 33 A continuous
Coil Spacings (centered) 300 mm
Max. Magn. Field Strength 2100 A/m (5 min.)
Nominal Magn. Field Strength 1260 A/m continuous
Magn. Field Strength, 1 A Coil Current 38.17 A/m; 151.63 dBµA/m
Coil Diameter 600 mm
Max. Cube-shaped DuT 32.5 x 32.5 x 32.5 cm
Mechanical Dimensions 0.64 m x 0.79 m x 0.42 m
Terminals 4 mm female
Inductance (Single Coil) 340 µH
Inductance (Coil Pair) 750 µH
Usable Frequency Range DC - 500 kHz
Resonance Frequency > 700 kHz
Weight 20.4 kg
Reference spacing for field values 300 mm

Where specified, coil distances refer to the coil centers. Continuous and short-duration limits are listed separately. Standards references retain the designations used in the source documents.

Installation and operating safety

Read the supplied datasheet before use. Switch off the complete system before making or removing connections and do not touch terminals during operation. Observe all voltage, current and time limits. High currents heat the coils; maintain adequate air circulation. Keep unintended magnetic field sources and magnetic materials away from the test volume.