FESP 5410-1, DC - 10 MHz, 1 kA, Square Induction Coil

  • Square induction coil
  • DC - 10 MHz, 400A Cont. 600A 10min
  • up to 1200A/m 5sec with MFPO 9760
  • 1-meter square
  • turns 1
  • Immunity testing to IEC 61000-4-8, IEC 61000-4-9 and IEC 61000-4-10

MARKETS:

Data Sheet

SPECIFICATIONS

 

FESP 5410-1- Square Induction Coil

The square induction coil FESP 5410-1 is used to generate magnetic fields for immunity testing according to IEC 61000-4-8, IEC 61000-4-9, and IEC 61000-4-10. The side length of the square-shaped coil is 1 m. The coil comes with screw terminals.

Operation as Helmholtz Coil:

Two square induction coil FESP 5210-1 can be combined to form a Helmholtz coil pair. The main advantages of a Helmholtz coil pair are an increase in field strength and a significant improvement of field uniformity. For this purpose, the coil planes are aligned in parallel. Electrically the coils have to be wired in series, keeping in mind that the current through the coils must flow in the same direction. This is very important to obtain the wanted addition of both field strength contributions in the volume between the coils.

 

Specifications:   

  Frequency range: DC ... 10 MHz
  Edge length: 1 m +/- 1 %
  Number of turns: 1
  Max. cont. current: 400 A
 

Max. current (10 min):

600 A
  Max. current (5 sec): 1000 A
  Max. cont. field strength
(coil center):
370 A/m
 

Inductance:

2.9 µH
     
  Material / diameter: Aluminum 10 x 25 mm
  Mounting tube: 22 x 195 mm
  Weight: 4.1 kg
  Connector: M16 Threads
 

Standard:

 

IEC 61000-4-8
IEC 61000-4-9
IEC 61000-4-10

 

  

 

  

 

 

 

OTHER PRODUCTS IN THE SAME CATEGORY:
Loading...