EM Clamp for immunity tests IEC / EN 61000-4-6 Ed. 4.0 • EM Clamp for immunity testing of cables with up to 20 (35) mm diameter • High coupling factor: less than 15 watts amplifier output power is required to obtain a test level of 10 V • Calibration unit and calibration data are supplied with each unit.
PARTNER:
MARKETS:
TEST STANDARDS:
CDN EMCL-20 & CDN EMCL-35
|
EM Clamp for immunity tests IEC / EN 61000-4-6 Ed. 4.0
|
· The preferred coupling and decoupling devices are the CDNs, for reasons of test reproducibility and protection of the AE. However, if they are not suitable or available, clamp injection can be used. · Often, clamp injection needs to be applied to multi-pair balanced cables because suitable CDNs might not be available. · The EM Clamp establishes both capacitive and inductive coupling to the cable connected to the EUT.
|
The EM-Clamp (in contrast to the conventional current injection clamp) has a directivity ≥ 10 dB, above 10 MHz, so that defined impedance between the common-mode point of the AE and the ground reference plane is no longer required. Above 10 MHz, the behavior of the EM Clamp is similar to that of a CDN.
|
EMCL-20 clamp for frequency range 10 kHz - 150 kHz
The EM clamp is developed for test frequencies above 150 kHz. For product standards with testing requirements in the frequency range of 10 kHz – 150 kHz the EMCL-20 clamp cannot operate like normal with a 6dB attenuator. The insertion loss for an effective coupling is too high. Therefore, a matching network CDN EMCL-NW_10 should be used. This network transforms the low impedance of the EM clamp to an acceptable impedance for the amplifier. With this matching network reproducible tests are feasible till 10V at 80% AM – without the 6 dB attenuator.
The matching network CDN EMCL-NW_10 is not required for the EMCL-35. EMCL-20 clamp for frequency range 10 kHz - 150 kHz
|
Specifications |
EMCL-20 |
EMCL-35 |
Frequency range |
10 kHz - 1000 MHz |
|
Nominal impedance |
50 Ohm |
|
Connector |
N-type, female |
|
Max. input level
0,15 – 100 MHz 100 – 230 MHz 230 – 1000 MHz |
100 W, 15 min 100 W, 5 min 50 W, 3 min |
|
Cable diameter |
< 20 mm |
< 35 mm |
Dimensions(L x W x H) in mm |
632 x 120 x 80 |
666 x 135 x 120 |
Weight |
7 kg |
14 kg |
Calibration unit of the CDN-EMCL-20 (included in scope of delivery of the EMCL -20 and EMCL-35)
|
Accessories (Calibration unit) |
|
Mounting bracket incl. 50/150 Ohm adapter (2 pieces) |
|
BNC termination, 50 Ohm, 1 W |
|
Adapter with 2 banana-plugs, length 34 mm |
For calibration of the network-analyzer |
Brass rod 4 mm with banana-plugs, length 672 mm |
For setting of test level |
Product information |
|
CDN-EMCL-20
CDN EMCL-NW_10 |
EM-Clamp, acc. IEC/EN 61000-4-6 for cables up to 20 mm diameter, incl. calibration unit and calibration data. Matching network 10 kHz – 150 kHz, for using EMCL-20 from 10 kHz to 150 kHz |
CDN-EMCL-35 CDN ABCL-20 |
EM-Clamp, acc. IEC/EN 61000-4-6 for cables up to 35 mm diameter, incl. calibration unit and calibration data. Decoupling clamp for additional decoupling at immunity testing according to IEC / EN 61000-4-6, 100 kHz – 1000 MHz, cable diameter max. 20mm. |