IGUU 2916, (Discontinued) Calibration-Pulse Generator for CISPR 16 Receivers

  • Universal Calibration Pulse Generator
  • acc. CISPR 16 for Bands A, B, C, D (9 kHz - 1000 MHz)
  • especially for EMI Receiver Tests (Pulse weighting, Overload)
  • Signal generator 100 kHz, 1 MHz, 10 MHz, 100 MHz (60 dBµV)
  • Pulse Repetition Frequency 0.1 Hz - 200 Hz (20 kHz)
  • Output Level adjustable in 1 dB steps
  • GPIB-Interface.

MARKETS:

TEST STANDARDS:

Differences between IGUU 2916 & 2918

KU 9618 Coax switch Option

Software Option

Manual

Data Sheet

SPECIFICATIONS

 

IGUU 2916, (Discontinued) Calibration-Pulse Generator for CISPR 16 Receivers


See IGUU 2918
  1. The IGUU 2916 covers the complete frequency range from 9 kHz - 1 GHz. Obviously, both relay contact generators with their 1-dB amplitude setting (main generators) are included with slight modification, because there is still no alternative.
  2. New Auxiliary Generators open the pulse frequency range beyond 200 Hz. So finally this limit of the mechanical relay contact has been overcome. Pulse frequency in the range of 0.1 Hz to 20 kHz is now available.
    • This means complete testing of emi-receivers including detectors for cispr-quasi-peak, average, and peak/mil.
  3. Rotary encoders and led-displays are used for pulse frequency and amplitude (main generator only) tuning and reading. Pulse frequency generation uses a crystal reference generator and dividers.
  4. The Sine Wave Generator provides the frequencies 100 kHz, 1 MHz, 10 MHz, and 100 MHz with 60 dBmV. The generator uses output voltage control and is referenced to the crystal generator using frequency dividers and multipliers.
    • This means sine wave/pulse comparison without an external signal generator.
  5. The optional Coaxial Switching Unit KU 9616 connects one of the four outputs of the IGUU 2916 to the e. u. t., controlled by the IGUU 2916.
    • This is very convenient and time-saving because there is no need to change coaxial connections.
  6. The optional IEEE-Interface serves for complete PC-control of all generator functions.
    • Automatic testing is possible by combining the interface and coaxial switching unit.

 

1. Main Generators
  • Pulse frequency range:
    • 0.1 Hz-200 Hz 0-20 Hz in 0,1 Hz-Steps,
    • 0 -200 Hz in 1 Hz-steps
  • Pulse frequency accuracy: better +/ -2x10 -4
  • Pulse shape: Almost rectangular
  • Pulse duration:
    • Band A: 270 ns,
    • Band B: 6 ns,
    • Band C/D: 0,3ns
  • Pulse polarity: positive/negative (switchable)
  • Pulse level: (CISPR weighting Q.P.):
    • Band A/B 10 -70 dB, in 1 dB-steps,
    • Band C/D 0 -60 dB, in 1 dB-steps
  • Max. Output Voltage acr. 50 W : 80 V - 110 V
  • Pulse Area at 60 dB attenuator setting: Corresponds to CISPR 16 Standard equivalent to 1 mVrms across 50 W with standardized Q.P.-Detector.
E*t Band A 9kHz - 150kHz 13.5 µVs
E*t Band B 150kHz - 30MHz 0.316 µVs
E*t Band C / D 30MHz - 1000MHz 0.044 µVs
  • Frequency flatness:
    • + / - 0dB to - 0.5dB at upper frequency limit for Band A / B
    • Band C: typ. + / - 0,5dB 30 - 100MHz (100MHz is reference frequency for level calibration).
    • Level increase between 200MHz and 800MHz + 0,5 to + 1dB, 800 - 1000MHz decrease to + / - 0dB to - 1dB.
  • LEVEL ACCURACY:
    • Band A / B: + / - 0,5dB at 50kHz, 1MHz
    • Band C / D: + / - 0,8dB at 100MHz.
2.Auxiliary generators
  • Pulse frequency ranges: 0.1 Hz-20 Hz, 1 -200 Hz, 10 Hz-2 kHz, 100 Hz -20 kHz.
  • Steps: 0,1 Hz, 1 Hz, 10 Hz, 100 Hz
  • Pulse frequency accuracy: better +/ -2x10 -4 for 0-200 Hz better +/ -2.10 -3 for 0-2 kHz, +/ -1,5% for 0 -20 kHz
  • Pulse duration typ.: BandA:70 ns, Band B:5 ns, Band C/(D):1 ns
  • Pulse level: ((CISPR weighting Q.P.): Band A/B/C/(D) 40 dB fixed level
  • Frequency flatness: Band A/B same as 1 st Main Gen. Band C(D): typ. +/ -0,75 dB 30 - 300 MHz -3 dB >500 MH
3. Sine Wave Generators
  • Frequencies: 100 kHz/1 MHz/10 MHz/100 MHz +/ -10-5
  • Output Level: 60 dB mV +/-0.5 dB
4. Dimensions: W x H x D 447 mm x 180 mm x 460 mm

 Options:

Opt. RecTest Softw.

Receiver Test Software for IGUU 2916 or 2918 signal generator and EMI-receiver to perform automatic calibration of an EMI receiver acc. to CISPR 16-1-1.

 

Opt. KU 9618

Coaxial Switching Unit for automatic performance tests with IGUU 2918, N-Connectors female.

 

 