30kHz - 1MHz
Use with FESP 5132
Per IEC 61000-4-39
PARTNER:
MARKETS:
CATEGORIES:
TEST STANDARDS:
SY 9223-439 TransformerDescription The SY 9223-439 transformer causes a voltage transformation Uin / Uout = 1 / 2 and is designed for operation with hand-held field coil – e.g. FESP 5132 when performing tests according to EC 61000-4-39. For this purpose, the field coil must be excited with the signal voltage of over 60 Vrms at 134 kHz to reach the field strength of 65 A/m.
Other applications in the frequency range up to 1... 3 MHz are possible.
The signal source is a signal generator connected to an audio amplifier with very low output resistance (e.g. LFPA 9733).
The SY 9223-439 reduces the requirements for audio amplifiers. The amplifier with the transformer connected to its output only needs to reach half the voltage required for the field coil.