HHS 5230-100, DC - 5 kHz, 3m Square Helmholtz Coils

Square Helmholtz Coils

Precisely defined magnetic fields. DC - 5 kHz.

The Schwarzbeck HHS 5230-100 uses 3 m square coils for magnetic immunity testing of whole vehicles. Designed for DC to 5 kHz, it supports applications described in SAE J551-17.

MARKETS:

TEST STANDARDS:

Data Sheet

Assembly Instructions

SPECIFICATIONS
DC - 5 kHzUsable frequency range
8 AContinuous nominal coil current
100 turnsPer coil
Defined by geometry

A calculable field.
A traceable measurement.

Magnetic field strength is proportional to coil current. Coil geometry, the number of turns and the applied current determine the resulting field, allowing the field measurement to be traced to a current measurement or the voltage drop across a known resistor.

The field depends on coil spacing and the position within the test volume. Use the reference spacings in the specifications and the field distribution data in the datasheet when selecting your setup.

Plan your test setup

Match the field to your application.

Continuous operation

The nominal coil current is 8 A continuous. The specified nominal magnetic field is 325 A/m @ coil Dist. = 1.8 m. Field values apply at the stated reference spacing.

Short-duration operation

The maximum coil current is 16 A (5 min.). Observe the stated time limit and provide adequate ventilation. The frequency range alone does not imply that maximum current or field is available at every frequency.

Technical details

Specifications

Product specifications with reference conditions. See the linked manufacturer datasheet for complete field distributions and operating guidance.

View the complete datasheet →
Number of turns 100
Maximum Coil Current 16 A (5 min.)
Nominal Coil Current 8 A continuous
Max. Magn. Field Strength 650 A/m @ coil Dist. =1.8 m
Nominal Magn. Field Strength 325 A/m @ coil Dist. = 1.8 m
Magn. Field Strength, 1 A Coil Current 50.8 A/m (coil Dist. 1.13 m)
Maximum Coil Separation >1.13 m
Coil Separation for best uniformity 1.8 m
Mechanical Dimensions W:2.0 m x H:3.10 m x B:3.18 m
Ground Clearance 0.12 m
Usable Frequency Range DC - 5 kHz
Resonance Frequency (per Coil) >10 kHz, typ. 13 kHz
Weight (per Coil) 190 kg
Recommended Accessories NFCN 9731-100  

Where specified, coil distances refer to the coil centers. Continuous and short-duration limits are listed separately. Standards references retain the designations used in the source documents.

Supporting documents

Installation and operating safety

Read the supplied datasheet before use. Switch off the complete system before making or removing connections and do not touch terminals during operation. Observe all voltage, current and time limits. High currents heat the coils; maintain adequate air circulation. Keep unintended magnetic field sources and magnetic materials away from the test volume.