Continuous operation
The nominal coil current is 9 A continuous. The specified nominal magnetic field is 1310 A/m continuous. Field values apply at the stated reference spacing.
Precisely defined magnetic fields. DC - 10 kHz.
The Schwarzbeck HHS 5210-100 generates defined magnetic fields from DC to 10 kHz. Its 100-turn coils provide a calculable relationship between drive current and magnetic field strength.
DC - 10 kHz
Magnetic field strength is proportional to coil current. Coil geometry, the number of turns and the applied current determine the resulting field, allowing the field measurement to be traced to a current measurement or the voltage drop across a known resistor.
The field depends on coil spacing and the position within the test volume. Use the reference spacings in the specifications and the field distribution data in the datasheet when selecting your setup.
The nominal coil current is 9 A continuous. The specified nominal magnetic field is 1310 A/m continuous. Field values apply at the stated reference spacing.
The maximum coil current is 15 A (5 min.). Observe the stated time limit and provide adequate ventilation. The frequency range alone does not imply that maximum current or field is available at every frequency.
Product specifications with reference conditions. See the linked manufacturer datasheet for complete field distributions and operating guidance.
View the complete datasheet →| Number of turns | 100 |
|---|---|
| Maximum Coil Current | 15 A (5 min.) |
| Nominal Coil Current | 9 A continuous |
| Inner side length of the coil | 1.0 m |
| Electrical side length of the coil | 1.04 m |
| Maximum Coil Separation | 0.70 m |
| Usable Frequency Range | DC - 10 kHz |
| Inductance (per Coil) | 31.3 mH |
| Resonance Frequency (Pair of Coils) | >25 kHz |
| Weight | 36.5 kg |
| Terminals | 4 mm female |
| Coil Separation acc. IEC 1000-4-8 | 0.40 m |
| Max. Magn. Field Strength | 2183 A/m (5 min.) |
| Nominal Magn. Field Strength | 1310 A/m continuous |
| Magn. Field Strength, 1 A Coil Current | 145.54 A/m (Coil Dist. 0.4 m) |
| Coil Separation for best uniformity along axes through the coil center | 0.566 m |
| Magn. Field Strength, 1 A Coil Current | 124.66 A/m (Coil Dist. 0.566 m) |
| Recommended Accessories | NFCN 9732-120 |
| Nominal/maximum field reference spacing | 0.40 m |
| Mechanical Dimensions | 1.15 m × 1.20 m × 0.75 m |
Where specified, coil distances refer to the coil centers. Continuous and short-duration limits are listed separately. Standards references retain the designations used in the source documents.
The compensation network NFCN 9732-120 consists of a capacitor and a fuse in series. It is used for series compensation of the HHS 5210-100 at electrical mains frequencies of 50 Hz to 60 Hz.
The compensation capacitor reduces the total impedance of the series circuitry consisting of the Helmholtz coils HHS 5210-100 and the NFCN 9732-120 at the operating frequencies of 50 to 60 Hz. It allows continuous currents of up to 8 A rms at generator voltages of less than 70 Vrms ...80 Vrms
Read the supplied datasheet before use. Switch off the complete system before making or removing connections and do not touch terminals during operation. Observe all voltage, current and time limits. High currents heat the coils; maintain adequate air circulation. Keep unintended magnetic field sources and magnetic materials away from the test volume.
Contact Absolute EMC LLC for the HHS 5210-100 in the United States, Mexico and Canada.